|
email
Comparison of EPR Oximetry with Two
Commonly Used Methods for Assessment of Tissue Oxygenation
during Steady State
and Respiratory Challenge
Julia O’Hara1, Huagang Hou1, Eugene Demidenko2,
Nadeem Khan1, Hongsheng Yu1, Roger Springett1,
Harold M. Swartz1
1EPR Center for the Study of Viable Systems,
Department of Diagnostic Radiology,
Dartmouth Medical School, Hanover, NH
03755, USA;
2Section of Biostatistics and Epidemiology,
Dartmouth-Hitchcock Medical Center, Lebanon, NH 03756, USA
INTRODUCTION
The goal of this study
is to provide comparisons of EPR oximetry with methods commonly in use or
currently under development for measurement
of tissue oxygenation. In this
study we compared measurements of rat brain cortex pO2 obtained
using EPR oximetry with those made using
either the Eppendorf microelectrode or
the Oxylite fluorescence based system. To optimize the utility of the data
obtained, we used each method
in the manner in which it is commonly used. That
is, for EPR oxygen sensitive material (LiPc) was implanted and then a period of
healing allowed
before making measurements. This is possible because
non-invasive measures of pO2 can be made after the implantation. To
optimize the potential
for good agreement and to fully utilize the capability
of EPR oximetry to make non-invasive measurements once the material is
implanted,
we made measurements with EPR and Eppendorf or Oxylite in the same
animal. In the initial study EPR and Eppendorf measurements
were made during normoxia, with sequential measurements. We also made simultaneous EPR and
Eppendorf as well as EPR and Oxylite
measurements during normoxia and a change
in FiO2 to 95%O2:5%CO2.
METHODS
Animals: Fischer CDF, 200-225g, male, artificially
ventilated and cannulated for blood gas sampling. All measurements were made
under normal physiologic
parameters of MABP, HR, temperature, paO2,
paCO2, and blood saturation.
EPR Oximetry:
A chronic implant of oxygen sensitive
material (LiPc, <100mg) was stereotactically
placed on the left side of the brain of rats
at a depth of about 1.5-2 mm below
the dura via a 25ga needle equipped with a plunger. Measurements
were made using an L-band (1.2GHz)
spectrometer using a loop resonator placed
above the skull. Care was taken to avoid artificial line broadening. In most
cases,
single scans of 7s were used. Linewidths were converted to pO2
using a calibration curve for the specific LiPc used.
Eppendorf pO2 Histograph 6650:
The Ag/AgCl reference
electrode was attached to the ear of the rat. The calibrated needle electrode
(0.27mm dia)
was inserted via a burr hole in the skull and moved from the dura
vertically into the cortex, taking measurements at 0.5 mm intervals to a depth of
4.5 mm.
Track 1 was on the left side, adjacent to the LiPc implant. Tracks 2
and 3 were on the contralateral side. The time period for the Eppendorf
measurements were about the same as that required for the EPR measurements
(total about 10 minutes), which includes setting up of the animal.
Oxylite:
The Oxylite system depends on the oxygen
sensitivity of a chromophore fixed to the end of a fiber designed to remain in
place
and provide measurements over time. In this case the EPR and Oxylite
measurements were made simultaneously on the same side of the brain.
We
compared both absolute tissue pO2 during normoxia and hyperoxia, as
well as relative pO2 during a change in FiO2 from
normoxia (26%O2)
to elevated oxygen using carbogen (95% O2:
5%CO2).
Data Analysis: An ANOVA model was used to analyze the
variance of Eppendorf and EPR techniques and to determine the relationship
between the methods. T-tests were used to compare means of values between Oxylite and EPR measurements. Significance level was p <0.05.
RESULTS
A comparison of absolute
Eppendorf and EPR measurements demonstrated: 1) That when comparing values
averaged over the groups of animals,
the two methods reported similar but
statistically different values with the EPR method reporting a somewhat higher
average pO2; 2)
There was poor correlation between the values
measured with the two techniques in the same animal on the same side of the
brain; 3)
The Eppendorf method reported a larger range of values of point pO2;
4) Consideration of the distribution of the heterogeneity of oxygen levels
in
the brain and the volumes sampled by the two methods provide an adequate
explanation for most of the observed differences; 5)
The differences in the
average pO2 measured by the two methods are consistent with other
studies that suggest that the Eppendorf electrode may,
under some
circumstances, report a depressed pO2, perhaps due to local
perturbations from the method.
Results from the
sequential assessment of brain pO2 with Eppendorf and EPR in the
same brain are shown in Figure 1. The density of Eppendorf
values is derived
from analysis of the frequency histogram of tissue PtO2 values,
using local smoothing by the Gaussian kernel with the default window
(S plus
6.1, Insightful, Seattle, WA) (1). For example, the
probability of having values of PtO2 from 10 to 30 is the area under
the density/curve,
approximately 0.02*(30-10) = 40%.
|

|

|
|
Figure
1: Distribution of Brain PtO2 values measured by Eppendorf and EPR
Oximetry. These
data include all Eppendorf values regardless of location. The EPR values for
each rat are represented by the vertical lines. (n=10rats)
|
Figure 2: Brain Cortex
PtO2 Using Oxylite and EPR Oximetry. Values were averaged
for 10 min under each condition. (n=7rats)
|
The method for use of
the Oxylite allows for placement of the probe in one site and continuous
monitoring of brain temperature and PtO2.
We established a routine
of taking a baseline measurement of cortex PtO2 with EPR and then
stereotactically implanting the tip of the Oxylite optode
in proximity to the
LiPc. Oxylite readings were initially very low compared to later values and to
baseline EPR made before optode insertion
and rose steadily after insertion of
the optode, requiring 30-60 minutes for stabilization. Comparisons were made
during 10 minutes intervals 1)
after stabilization of the PtO2, 2)
while the rat was breathing carbogen (followed for 15-25 minutes after changing
breathing gas) and 3) during recovery,
about 30-40 minutes after switching back
to normoxic conditions. Both methods showed similar absolute values and changes
in brain PtO2.
Results are shown in Figure 2 above. There was a
difference in baseline (p =0.0002), but there was no difference during carbogen
or during recovery period.
In 9 rats tested, all responded to the changes in
FiO2 with rapid and consistent increases in tissue PtO2.
In 2 of the animals,
the Oxylite signal became unstable and therefore no value
could be obtained for PtO2 during the carbogen period.
Therefore,
these animals were excluded from analysis.
CONCLUSION
These results indicate
that the three methods provide different views of oxygen in tissue, and
therefore it is difficult and probably not usefu
to attempt to characterize
one of these as “better”. Overall they reported similar values for average PtO2
in the cortex of rat brain.
They differ in the volumes that are sampled in a
single measurement, the number of measurements, and the occurrence of trauma
associated
with the measurement. The Eppendorf histograph reveals tissue
heterogeneity that cannot be resolved by EPR oximetry or the Oxylite
because of
the smaller volume that is a sampled by the Eppendorf. The Eppendorf and Oxylite
techniques inherently can cause trauma
at the time that the measurement is
made. The EPR method provides a means to make repeated measurements without
closely
associated trauma but does not resolve small regions of heterogeneity.
REFERENCE
(1). Silverman, B. W. (1986). Density Estimation
for Statistics and Data Analysis. Chapman and Hall, London.
ACKNOWLEDGEMENTS
This work
was supported by a NIH (NIBIB) grant PO1 EB002180, “Measurement of pO2 in
Tissues In Vivo and In Vitro,”
and used the facilities of the EPR Center for the Study of Viable
Systems supported by NIH (NIBIB) grant P41 EB002032.
back to research list
home
|